Autoclaves are sterilizers using high pressure and high temperature steam to sterilize media, glassware, instruments, waste, etc. To accomplish the desired end goal — and to protect the user and the environment from hazardous materials — the autoclave must be used correctly. Additionally, wastes must be managed in compliance with state and local regulations.
Physical hazards involved with steam autoclaves are heat, steam and pressure. The biological hazards involve potential exposure to viable human pathogens.
Due to the high heat and pressure created in autoclaves during operation, you must follow proper loading, use and unloading procedures to prevent burns and other accidents. Burns can result from physical contact with the structure of the autoclave, and steam burns can occur from contact with steam leaving the apparatus. Burns can also result from careless handling of vessels containing hot liquids. Explosive breakage of glass vessels during opening and unloading — as a result of temperature stresses — can lead to mechanical injury, cuts and burns.
Autoclave performance for sterilization purposes is dependent on proper use.
If you are a principal investigator (PI), it is your responsibility to ensure that:
Personnel who use an autoclave must be trained to understand the proper operation procedures as referenced in:
Video training is also available: “On-Site Biohazardous Waste Management: Autoclaving.”
Additionally, instruction sheets are posted in each autoclave room to assist with the operation of the equipment.
You must autoclave all potentially infectious materials before you wash, store or dispose of them.
You must label biohazardous materials as such, and you must sterilize them by the end of each work day, or you must secure them appropriately. Biohazardous materials should not be left in an autoclave overnight in anticipation of autoclaving the next day.
For the autoclave process to be effective in achieving sterilization, sufficient temperature, time and direct steam contact are essential. Air must be completely removed from the sterilizer chamber and from the materials to allow proper steam penetration.
Factors that affect air removal include:
The potential safety risks for autoclave operators are:
You must wear personal protective clothing and equipment when loading and unloading the autoclave, including:
Ensure that the material is able to be autoclaved. Items that should not be autoclaved include:
Glassware should be heat-resistant borosilicate (Pyrex or Kimax), and you should inspect it for cracks prior to autoclaving.
Plastics should be heat resistant, for example:
Prepare and package material suitably. You should wrap or bag loose dry materials in steam-penetrable paper or loosely cover them with aluminum foil. Wrapping too tightly will impede steam penetration, decreasing the efficiency of the process.
Containers of liquid should be a maximum volume of 2/3 filled.
Cover all containers with a loosened lid or steam-penetrable bung to prevent pressure buildup and to prevent bottles from shattering during pressurization.
Use plain, unmarked containers for items that are not hazardous.
You should tag items or baskets with autoclave temperature tape.
Place items in containers to secure and contain spills. Place the following items inside a secondary pan (must be autoclavable plastic or a stainless-steel container) in the autoclave:
The pan must be large enough to contain a total spill of the contents. Open, shallow, metal pans are more effective in conducting heat and allowing air removal than tall plastic tubs. Adding some water to the secondary pan will help to heat items more evenly.
The parameters for the sterilization cycle will depend upon the amount and type of material. Usually 121 degrees Celsius at 15 pounds-force per square inch (lbf/in2 or psi) for a minimum of 30 minutes is recommended. However, you can increase the temperature and cycle time depending on the size and type of load.
Autoclave manufacturers generally provide several pre-set cycle options that vary in the pre-set sterilization temperature, sterilization time and dry time.
After a run is complete, check the pressure gauge to ensure that the pressure in the chamber is “0.” If pressure is not released, do not open the door. Inform your PI / supervisor about the malfunction.
Before opening the door, wear eye protection and heat-resistant gloves or mitts. Be sure to wear closed-toed shoes (hot condensate may drip from the door). Rubber aprons in addition to rubber sleeve protectors are advisable.
Use caution when removing liquids, molten agar, etc. Liquids, especially large volumes, may continue boiling for some time after autoclaving. To avoid being splashed with scalding liquid:
Slide a cart to the opening of the autoclave and pull the autoclave secondary container onto the cart for transport. Place the cart in a low-traffic area while additional cooling occurs.
Before touching items with ungloved hands, allow for the following cooling times:
Please note: If a faulty condition exists (e.g., sterilizer did not finish the cycle, or water leaks out when the door is unlocked), inform the person responsible for contacting the autoclave service technician.
When you are autoclaving biohazardous waste materials inside heat-resistant autoclavable bags, be sure that you:
It is advisable to add some water to bags of solid wastes (the water will vaporize into steam that will drive out residual air once sterilization temperature has been reached inside the bag).
Label the autoclavable bag with a “biohazard” symbol on it with commercially available autoclave temperature tape that changes color, e.g., visible black stripes or the word “autoclaved” appear once sterilizing temperature has been reached. Autoclave temperature tape only indicates that the desired temperature was reached — it does not indicate any information about time and pressure. Apply this tape across the “biohazard” symbol on the bag before autoclaving. Do not throw away any biohazard bags without covering all biohazard symbols.
Autoclave 50 to 60 minutes, at temperature and pressure of 121 degrees Celsius (250 degrees Fahrenheit) and 15 psi. When cycle is finished, inspect autoclave temperature tape and visually check autoclaved bags. Bags should be left to cool for several minutes before removing from autoclave.
Once cool, remove the Steam Chemical Integrator from the autoclave bag and interpret the results to ensure that the autoclave worked correctly. If this test indicates a successful sterilization, then:
Please note:
Document the treatment of each load of biohazardous waste on the Autoclave Waste Treatment Log (available at each equipment location), which lists:
Keep printout strips with the log sheets as documentation of the autoclave operation.
No person shall operate an autoclave unless it has been inspected by a qualified inspector and a certificate of inspection has been issued. A current inspection certificate is posted near the autoclave.
Users are not to make repairs. Autoclaves shall be maintained and repaired by qualified persons.
If the autoclave does not operate exactly as expected, Facilities Planning and Management will place a notice on the autoclave indicating that it is not to be used until the problem is diagnosed and corrected.
Facilities will record the problem in autoclave log book.
When maintenance work or repairs are needed, the PI/LAB must provide a safe work environment for the service technician. Please remove all items from the sterilizer chamber, clean any spills or leaks inside the chamber, remove untreated biohazardous materials from the vicinity, etc.
If any injury occurs, seek first aid or, if necessary, seek medical assistance by dialing:
If clothing is soaked in hot water/steam, remove clothing and place the injured area in cool water.
Place a notice on the autoclave indicating that it is not to be used until:
Report all other non-emergency incidents to Facilities Planning and Management.
Spills may occur from a boil-over or breakage of containers. You are not allowed to operate the autoclave until the spill is cleaned.
If you are the operator, you are responsible for cleaning spills. The spilled material can be contained using materials from the spill kit to absorb or contain the spill.
Wait until the autoclave and materials have cooled to room temperature before starting the cleanup inside the autoclave.
Review the Safety Data Sheet(s) (SDS) of the spilled material(s), if appropriate, to determine which protective equipment, spill cleanup, and disposal protocols are necessary.
You must:
If materials have been intermingled, follow the cleanup and disposal protocol for the most hazardous component of the mixture.
Cracked clean glassware must be disposed of properly in the “broken glass” disposal container provided by Custodial Services / EH&S.
Record the spill and cleanup procedure in autoclave log book.
Inform Environment, Health and Safety so that the spill kit can be replenished.
Proper personal protective equipment (PPE) must be worn when replacing the cleaning chemicals.
To achieve maximum results, the Auto Clave Glassware Washers (Lancer 1400LX) require two separate chemical items (Consumables) to function properly:
These two chemicals have been primarily used at the Jacobs School.
For the Lancer 1400LX: 2 (two) glassware washers per each autoclave room. (The 2.5-gallon container is required of LCD-P and LCA-A for each machine) – so this would be one box of each that contains 5 gallons total of each product as listed here:
The Lancer Clean Detergent is formulated and is specifically designed for our Lancer high-pressure washers to achieve maximum cleaning results. The solution is powerful enough to clean a variety of contaminants, including laboratory soils, "baked on" organic residues and petroleum applications. This detergent is formulated to safely clean all types of glassware, surgical instruments and other labware items. Use of the chemical is made easy with their pre-mixed solution for direct use from the bottle to the washer and their peristaltic pump enabling automatic dosing – no measuring needed. The cleaning chemical is produced with a double filtration system and excellent quality control, ensuring the integrity of these solutions is maintained.
The Lancer Acid Rinse is to be used as a rinsing agent following the detergent in the wash cycle. The solution is specifically designed for our Lancer high-pressure washers to achieve maximum rinsing results. The solution is powerful enough to rinse a variety of contaminants including laboratory soils, "baked on" organic residues and petroleum applications. Use of this chemical is made easy with their premixed solution for direct use from the bottle to the washer and their peristaltic pump enabling automatic dosing – no measuring needed. This cleaning chemical is produced with a double filtration system and excellent quality control, ensuring the integrity of the solutions is maintained.
If the 2.5-gal container is empty (zero material remaining) it is safe to bag and dispose of the container according to the normal recycling procedures for the building.
If the container has some liquid remaining, you must complete the following tasks:
Environmental, Health and Safety will pick up the materials and complete proper disposal activities.